Publication:
Comparison Between Automated and Manual Sphygmomanometer for Measuring Blood Pressure in Children

creativeworkseries.issnISSN 1990-7974 eISSN 1990-7982
dc.contributor.authorTaksande, A
dc.contributor.authorJadhav, A
dc.contributor.authorVagha, J
dc.date.accessioned2026-03-22T09:59:43Z
dc.date.available2026-03-22T09:59:43Z
dc.date.issued2015
dc.descriptionA Taksande Professor, Department of Paediatrics, Jawaharlal Nehru Medical College A Jadhav Senior Resident, Department of Paediatrics, Jawaharlal Nehru Medical College J Vagha Professor and Head of Department, Department of Paediatrics, Jawaharlal Nehru Medical College
dc.description.abstractAbstract: Introduction: The mercury sphygmomanometer has been the gold standard used for obtaining blood pressure (BP). However, due to environmental concerns and more use of automated BP devices, an alternative to using the standard mercury sphygmomanometer to measure BP. The aim of this study is to evaluate the accuracy of the Automated (Omron) BP device against the mercury sphygmomanometer in children. Material and Methods: One hundred children were enrolled in this study. The accuracy of Omron HEM–7121 BP readings was compared with that of mercury BP device readings. Each children had four BP measurements recorded sequentially i.e. two valid systolic readings, and two valid diastolic readings. Results: BP taken with the automated device was systolic 109.77 ± 9.97 and diastolic 74.50 ± 8.32 mm Hg compared to systolic 112.68 ± 9.98 and diastolic 77.38 ± 7.91 mm Hg measured by manual mercury sphygmomanometer BP readings (p<0.001). Automated and mercury BP measurements were correlated (r = 0.89, systolic BP; r = 0.82, diastolic BP). Linear regression analysis showed that the automated systolic BP is a significant predictor (? = 0.897, p < 0.001) of manual systolic BP. Similarly, automated diastolic BP was also a significant predictor (? = 0.829, p< 0.001) of manual diastolic BP. Conclusion: This study concludes that the automated BP device is reliable and accurate for measuring the BP in children.
dc.identifierhttps://doi.org/10.3126/jnps.v35i1.11568
dc.identifier.urihttps://hdl.handle.net/20.500.14572/5323
dc.language.isoen_US
dc.publisherNepal Paediatric Society (JNPS)
dc.subjectAutomated devices
dc.subjectblood pressure
dc.subjectmeasurement
dc.subjectOMRON
dc.subjectMercury Sphygmomanometer
dc.titleComparison Between Automated and Manual Sphygmomanometer for Measuring Blood Pressure in Children
dc.typeArticle
dspace.entity.typePublication
local.article.typeOriginal Article
oaire.citation.endPage17
oaire.citation.startPage13
relation.isJournalIssueOfPublication5a1d626f-003d-4bd6-a558-b7a82329ac4e
relation.isJournalIssueOfPublication.latestForDiscovery5a1d626f-003d-4bd6-a558-b7a82329ac4e
relation.isJournalOfPublication6f9be05c-05a9-4a3e-a5b5-a19a15ab042c

Files

Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
13-17.pdf
Size:
308.48 KB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.86 KB
Format:
Item-specific license agreed to upon submission
Description:

Collections